4339B
Change 1
Delete
the
following
function.
Limit
V
alue
Reading
function
(
LimitC
)
Oset
Setting
function
(
OfsEnt
)
Dut
Capacitance Reading
function (
MeasC
)
Delete the
following GPIB
command.
:SENSe:CONTact:DATA?
:SENSe:CONTact:LIMit?
:SENSe:CONTact:OFFSet
Change
2
Change
T
able
C-1
and
T
able
C-2
to
T
able
A
-3
and
T
able
A
-4
respectively
.
T
able
A
-3.
O
VLD/Over-Current/N.C.
Operations
(Meas
.
parameter:
R)
Display
Handler Output GPIB Output
1
Solutions
Data
Mode
Comprtr
Mode
<
stat
>
2
:
1
O
VLD
O
VLD
LOW /LO
<
data
>
:
9.9E37
Select
an
appropriate
measurement
range
.
(Overload)
<
comp
>
3
:
4
/LO
<
stat
>
2
:
4
Reduce
the
source
voltage
.
Over-Current O
V
CURR
LOW &
<
data
>
:
9.9E37
Check
whether
a
short
circuit
is
occurred.
/NOT
READY
<
comp
>
3
:
4
<
stat
>
2
:
2
Cancel
the
bad
contact
between
DUT
and
N.C. N.C. N.C. /NO
CONT
A
CT
<
data
>
:
9.9E37
contact-pin.
(No-Contact)
<
comp
>
3
:8
O
VLD
<
stat
>
2
:
3
Select
an
appropriate measurement
range.
& N.C. N.C. /LO
<
data
>
:9.9E37
Cancel the bad contact between DUT and
N.C.
<
comp
>
3
:4 contact-pin.
OVLD /LO
<
stat
>
2
:5 Select an appropriate measurement range
.
& OVCURR LOW &
<
data
>
:9.9E37
Reduce the source voltage
.
Over-Current /NOT READY
<
comp
>
3
:4 Check whether a short circuit is occurred.
1
Refer to
:FETCh? in Chapter 5
2
Measurement status (0:Normal, 1:Overload, 2:No-Contact, 4:Over-Current)
3
Comparison result (1:In, 2:High, 4:Low, 8:No-Contact)