Calibration
Required
for
the Impedance
Analyzer Mode
Calibration Required
for the
Impedance Analyzer
Mode
This
section
provide
procedures
for performing
the calibration
when measuring
in the
impedance
analyzer
mode
.
This section
also covers
how to
customize the
user dened
calibration
kit.
OPEN/SHORT/LO
AD
Calibration
In
the
impedance
analyzer
mode
,
4395A
should
be
calibrated
with
the
43961A
impedance
test
kit
attached.
Calibration
denes
the
measurement
accuracy
at
the
OUTPUT
port
on
the
impedance
test
kit.
After
calibration,
the
analyzer
can
measure
within
its
specied
measurement
accuracy
.
The
calibration
is
performed
in
terms
of
the
three
items:
OPEN
(using
0
S
termination)
SHORT
(using
0
termination)
LO
AD
(using
50
termination)
Note
Y
ou must
use the
calibration
kit
that
is
furnished
to
the
43961A
for
the
standard
calibration
kit.
Calibration
Procedure
F
ollow
the
steps
below to
perform
the
OPEN/SHORT/LO
AD
calibrations
.
1.
Press
4
Cal
5
.
2.
Press
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
N
CALIBRATE
MENU
.
3.
Connect the
0
S
termination
to
the
OUTPUT
port.
4.
Press
N
N
N
N
N
N
N
N
N
N
NN
NN
OPEN
.
After
an
open
calibration
sequence
is
completed,
the
N
N
N
N
N
N
N
N
N
N
N
N
N
N
OPEN
softkey
label
is
underlined.
5.
Disconnect
the
0
S termination,
then connect
the 0
termination
to
the
OUTPUT
port.
6.
Press
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
N
SHORT
.
After
a
short calibration
sequence is
completed, the
N
N
N
N
N
N
N
N
NN
NN
NN
NN
N
SHORT
softkey
label is
underlined.
7. Disconnect the 0 termination, then connect the 50 termination.
8.
Press
NNNNNNNNNNNNNN
LOAD
.
After a load calibration sequence is
completed, the
NNNNNNNNNNNNN
N
LOAD
softkey label is underlined.
9.
Press
NNNNNNNN
NNNNNNNNNNNNNNNNNNNNN
DONE: CAL
.
10. V
erify the \
Cor
" notation is displayed on
the left of the screen.
7-14 Calibration