Calibration
Required
for
the Impedance
Analyzer Mode
Figure
7-1.
Connecting
Calibration
Standards
Note
The
OUTPUT
port
of
the impedance
test kit
and the
calibration
standards
have
APC-7 connectors
.
The
APC-7
connector
is
very
sensitive
to
damage
and
dirt.
You
need to
do
the
following
when
handling
and
storing
APC-7
connectors:
Keep the
connectors
clean.
Do
not
touch
the
mating
plane
surfaces
.
Do
not
set
the
connectors
contact-end
down.
Before
storing,
extend
the
sleeve
or
connector
nut.
Use
end
caps
over
the
mating
plane
surfaces
.
Never
store
connectors
loose
in
a
box
or
a
drawer
.
Connecting
the
T
est
Fixture
T
o
connect
the
test
xture
to
the
impedance
test
kit,
see
the applicable
test xture
manual for
instructions
.
The
following
is
a
general
procedure:
1.
Turn
the
APC-7
connector
of
the
impedance
test
kit
OUTPUT
port.
2.
V
erify
that
the
connector
sleeve
is
retracted
fully.
3. Set the
mounting posts of the test station into the twin locating holes at the corner of the
test xture
.
4. Connect the connector on
the underside of the test xture to the OUTPUT port of the
impedance test kit.
Calibration 7-15