53220A/53230A User’s Guide 115
53220A/53230A Measurements 3
MEASure:FREQuency:PRI? [{<expected> | MINimum | MAXimum |
DEFault} [,{<resolution> | MINimum | MAXimum | DEFault}]]
[,<channel>]
CONFigure:FREQuency:PRI [{<expected> | MINimum | MAXimum |
DEFault} [,{<resolution> | MINimum | MAXimum | DEFault}]]
[,<channel>]
- expected is the expected interval. resolution is the desired measure-
ment resolution in seconds. The parameter ranges are:
expected: 100 ns to 1.0s (default = 0.001s)
resolution: 1.0E-15 * expected to 1.0E-5 * expected
(default resolution corresponds to a 0.1s gate time)
- channel is counter channel 3 specified as (@3).
MEASure:FREQuency:PRF? [{<expected> | MINimum | MAXimum |
DEFault} [,{<resolution> | MINimum | MAXimum | DEFault}]]
[,<channel>]
CONFigure:FREQuency:PRF [{<expected> | MINimum | MAXimum |
DEFault} [,{<resolution> | MINimum | MAXimum | DEFault}]]
[,<channel>]
- expected is the expected repetition frequency. resolution is the desired
measurement resolution in hertz. The parameter ranges are:
expected: 1 Hz - 10 MHz (default = 1 kHz)
resolution: 1.0E-15 * expected to 1.0E-5 * expected
(default resolution corresponds to a 0.1s gate time)
- channel is counter channel 3 specified as (@3).