EasyManua.ls Logo

Agilent Technologies 53220A - Standard Event Register

Agilent Technologies 53220A
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
272 53220A/53230A User’s Guide
8 Instrument Status
Standard Event Register
The Standard Event Register monitors programming conditions including:
operation complete
query errors
device errors
execution errors
command execution errors
power-on state
Reading the Standard Event Register
The Standard Event Register is read with the command:
*ESR?
Reading the register does not clear the bits. The bits are cleared by sending
*CLS.
Enabling the Standard Event Register
The Standard Event Enable register specifies which bits in the Standard Event
Status register can generate a summary bit which is subsequently used to
generate a service request.
Bits within the standard event register are enabled with the command:
*ESE <enable>
*ESE? (query form)
- enable: decimal value corresponding to the binary-weighted sum of the bits
in the register. For example, to enable the bit monitoring command syntax
errors: *ESE 32.
*ESE? returns the weighted-sum of all enabled bits.

Table of Contents