118 53220A/53230A User’s Guide
3 53220A/53230A Measurements
// measure the negative (off) width of the burst signal
*RST // reset to start from known state
MEAS:NWID:BURS? (@3)
// measure the negative (off) width using CONFigure, set
// a -12 dB detector threshold
*RST // reset to start from known state
CONF:NWID:BURS (@3)
INP3:BURS:LEV -12 // set a detector threshold of -12 dB
READ?
Notes
1. See “Burst Measurement Detector Threshold” in Chapter 4 for information
on threshold settings and its relationship to positive and negative width mea-
surements.