178 53220A/53230A User’s Guide
5 Triggering and Gating
Figure 5-8. Gate Start Sequence.
Gate Start Source
The specific command which sets the gate start (open) source is:
Gate Delay
Gate Start Source
Gate Slope
ADVanced
IMMediate
EXTernal
SENSe:GATE:STARt:SOURce
SENSe:GATE:STARt:SLOPe
SENSe:GATE:STARt:DELay:SOURce
SENSe:GATE:STARt:DELay:EVENts
(input channel)
SENSe:GATE:STARt:DELay:TIME
IMMediate
Gate Start (open)
(internal)
*
not available with time interval
INPut[1] | INPut2 *
SENSe:GATE:EXTernal:SOURce
External Gate Source
BNC
(delay disabled)
Gate Source
SENSe:FREQuency:GATE:SOURce
SENSe:TOTalize:GATE:SOURce
SENSe:TINTerval:GATE:SOURce
(from Triggered State)
shown first sequentially
set last programmatically