53220A/53230A User’s Guide 183
Triggering and Gating 5
Figure 5-9. Gate Stop Hold Off and Gate Stop Sequence.
Gate Stop Hold Off
The gate stop hold off delays the gate from stopping (closing) for a specified
time or until a specified number of events (edges) have occurred on the input
channel.
For totalizing measurements or when specifying a hold off in terms of time,
the hold off begins when the gate opens. For all other measurements or when
specifying a hold off for a number of input events, the hold off begins on the
first event after the gate opens (Figure 5-1).
(from Gate Start (open))
*
not available with time interval
Gate Stop Source
SENSe:GATE:STOP:SOURce
IMMediate
Gate Slope
SENSe:GATE:STOP:SLOPe
Gate Stop (closed)
EXTernal
INPut[1]* | INPut2 *
SENSe:GATE:EXTernal:SOURce
External Gate Source
EXTernal /
(internal)
Gate Stop Hold Off
SENSe:GATE:STOP:HOLDoff:EVENts
(input channel)
SENSe:GATE:STOP:HOLDoff:SOURce
IMMediate
(holdoff disabled)
SENSe:GATE:STOP:HOLDoff:TIME
Gate stop source does not apply to rise/fall
time, duty cycle, pulse width, phase, and
single-period measurements