53220A/53230A User’s Guide 201
53220A/53230A Math, Graphing, and Data Logging 6
Assigning Reading Units
For ease in identifying readings from the front panel, a user-defined “units
string” can be assigned to any scale function. The user string replaces the
instrument-assigned measurement units (Hz, pct, ppm, ...). The unit prefix
( µ, m, k, M) remains.
The units string is displayed on the counter and defined using the commands:
CALCulate[1]:SCALe:UNIT:STATe {OFF|ON}
CALCulate[1]:SCALe:UNIT:STATe? (query form)
CALCulate[1]:SCALe:UNIT "<units>"
CALCulate[1]:SCALe:UNIT? (query form)
ON - displays user-defined units on the front panel. OFF - turns off the
user-defined units.