220 53220A/53230A User’s Guide
6 53220A/53230A Math, Graphing, and Data Logging
Resetting the Histogram
The data from which the current histogram is created is cleared by any of the
following:
pressing the Reset Histogram softkey
enabling/disabling or changing any part of the histogram
- number of bins, lower or upper range
a new measurement cycle - INITiate:IMMediate, READ?, MEASure
sending a new SCPI command or changing a current SCPI parameter
an instrument reset or present - *RST, SYSTem:PRESet
These actions also clear all readings in reading memory and in the output
buffer.
To clear histogram data without clearing the readings in reading memory, the
following command is used:
CALCulate2:TRANsform:HISTogram:CLEar[:IMMediate]