248 53220A/53230A User’s Guide
7 Formats and Data Flow
Figure 7-2. Folders and Files within Internal Memory and USB Storage.
root directory
sub-folder
folder
folder
MMEMory:MDIRectory “<folder
>”
(creates folders)
MMEMory:CDIRectory “<folder>”
(specifies current folder for
MMEMory Subsystem)
USB Storage
Device
Agilent 53220A/53230A Counter
MMEMory:STORe:DATA RDG_STORE, <file>
(creates data (.csv, .dat) files)
MMEMory:STORe:STATe <file>
(creates state (.sta) files)
file
file
file
file
U
SB host port
volatile reading memory
1M readings
battery
flash
instrument firmware
calibration
constants
state storage
counter flash (non-volatile) memory
internal flash
5M readings
USB:\
INT:\
Path is “Internal”
Path is External”