53220A/53230A User’s Guide 91
53220A/53230A Measurements 3
A time interval measurement is the difference between two events, or edges,
on different waveforms or on the same waveform. A typical two-channel time
interval measurement is shown in Figure 3-6.
Figure 3-6. 2-Channel Time Interval Measurement.
The commands used to make time interval measurements are:
MEASure:TINTerval? [<channel_pair>]
CONFigure:TINTerval [<channel(s)>]
- channel_pair are channels in the form: (@1),(@2)|(@2),(@1)
- channel(s)can be a channel pair or single channel in the form:
(@1),(@2) | (@2),(@1) | (@1)|(@2)
The input channel(s) are (default) configured for auto-leveling at 50% with a
positive slope.
When configuring time interval measurements consider the start or stop input
thresholds, input slope, or gate stop hold off settings: