EasyManuals Logo

Agilent Technologies 6460 Installation Guide

Agilent Technologies 6460
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #70 background imageLoading...
Page #70 background image
3 Installation Completion and System Verification
Step 5. Verify the Triple Quadrupole LC/MS sensitivity
70 Agilent 6400 Series Triple Quad LC/MS System Installation Guide
Tabl e 5 Installation Sensitivity Performance Specifications
Model Mode G1948B Electrospray Agilent Jet Stream
6495 Positive/Negative 75,000:1 (Auto-RMS,1pg)
6490 Positive/Negative 25,000:1 (Auto-RMS,1pg)
6470 Positive 37,500:1 (Auto-RMS,1pg)
Negative 15,000:1 (Auto-RMS,1pg)
6460 Positive 5,000:1 (Auto-RMS,1 pg) 15,000:1 (Auto-RMS,1 pg)
Negative 1,500:1 (Auto-RMS,1 pg) 5,000:1 (Auto-RMS,1 pg)
6430 Positive 300:1 (1×RMS, 1 pg)
6420 Positive 3,000:1 (Auto-RMS,1 pg)
Negative 1,000:1 (Auto-RMS,1 pg)
6410B
with
G2573A
Positive 300:1 (1×RMS, 1 pg)
6410B Positive 150:1 (1×RMS, 1 pg)
Tabl e 6 G6061A/G6062A/G6063A Marketing Bid Sensitivity Performance
Specifications
Model Mode G1948B Electrospray Agilent Jet Stream
6495 Positive/Negative 150,000:1 (Auto-RMS,1pg)
6490 Positive/Negative 50,000:1 (Auto-RMS,1pg)
6470 Positive (G6061A) 75,000:1 (Auto-RMS,1pg)
Negative (G6062A) 30,000:1 (Auto-RMS,1pg)
6460 Positive (G6061A) 10,000:1 (Auto-RMS, 1 pg) 30,000:1 (Auto-RMS,1 pg)
Negative (G6062A) 3,000:1 (Auto-RMS, 1 pg) 10,000:1 (Auto-RMS,1 pg)
6420 Positive (G6061A) 6,000:1 (Auto-RMS, 1 pg)
Negative (G6062A) 2,000:1 (Auto-RMS, 1 pg)

Table of Contents

Other manuals for Agilent Technologies 6460

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6460 and is the answer not in the manual?

Agilent Technologies 6460 Specifications

General IconGeneral
BrandAgilent Technologies
Model6460
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals