Language Dictionary - 8
123
INITiate:CONTinuous:SEQuence1
INITiate:CONTinuous:NAME TRANsient
These commands control the output transient trigger system.
1 or ON
continuously initiates the output trigger system..
0 or OFF
turns off continuous triggering. In this state, the output trigger system must be initiated
for each trigger using INITiate:SEQuence.
Command Syntax
INITiate:CONTinuous:SEQuence1 <bool>
INITiate:CONTinuous:NAME TRANsient <bool>
Parameters
0 | 1 | OFF | ON
Examples
INIT:CONT:SEQ1 ON INIT:CONT:NAME TRAN, 1
Related Commands
ABOR INIT TRIG TRIG:SEQ:DEF *TRG
TRIGger
This command generates a BUS trigger for the output transient trigger system. If the transient trigger
system is enabled, the trigger will then:
1. Initiate a pending level change as specified by CURRent:TRIGger or VOLTage;TRIGger.
2. Clear the WTG bit in the Status Operation Condition register after both transient and acquire trigger
sequences have completed. (WTG is the logical-or of both transient and acquire sequences.)
3. If INITiate:CONTinuous ON has been programmed, the trigger subsystem is immediately re-enabled
for subsequent triggers. As soon as it is cleared, the WTG bit is again set to 1.
Command Syntax
TRIGger[:SEQuence1][:IMMediate]
TRIGger[:TRANsient][:IMMediate]
Parameters
None
Examples
TRIG TRIG:IMM
Related Commands
ABOR CURR:TRIG INIT *TRG VOLT:TRIG
TRIGger:SOURce
This command selects the trigger source for the output transient trigger system. Since BUS is the only
trigger source for transient triggers, this command does not need to be used. It is included for
completeness only.
BUS
GPIB device, *TRG, or <GET> (Group Execute Trigger)
Command Syntax
TRIGger[:SEQuence1]:SOURce<source>
TRIGger[:TRANsient]:SOURce<source>
Parameters
BUS
*RST Value
BUS
Examples
TRIG:SOUR BUS
Query Syntax
TRIGger[:SEQuence1]:SOURce?
TRIGger[:TRANsient]:SOURce?
Returned Parameters
<CRD>

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 66321D and is the answer not in the manual?

Agilent Technologies 66321D Specifications

General IconGeneral
BrandAgilent Technologies
Model66321D
CategoryPower Supply
LanguageEnglish

Related product manuals