8 – Language Dictionary
126
TRIGger:SEQuence2:HYSTeresis:DVM
TRIGger:ACQuire:HYSTeresis:DVM
Agilent 66321D/66319D only
This command defines a band around the trigger level through which the input signal must pass before a
DVM measurement can occur. The band limit above and below the trigger level is one half of the
hysteresis value added to or subtracted from the trigger level. For a positive trigger to occur, the
excursion of an input signal in the positive direction must start below the lower hysteresis band limit and
pass through the upper hysteresis band limit. For a negative trigger to occur, the excursion of an input
signal in the negative direction must start above the upper hysteresis band limit and pass through the
lower hysteresis band limit.
Command Syntax
TRIGger:SEQuence2:HYSTeresis:DVM<NRf+>
TRIGger:ACQuire:HYSTeresis:DVM<NRf+>
Parameters
0 to MAX (see table A-3)
Unit
V (volts)
*RST Value
0
Examples
TRIG:SEQ2:HYST:DVM 0.5 TRIG:ACQ:HYST:DVM 0.5
Query Syntax
TRIGger:SEQuence2:HYSTeresis:DVM?
TRIGger:ACQuire:HYSTeresis:DVM?
Returned Parameters
<NR3>
Related Commands
TRIG:SEQ2:LEV:DVM
TRIGger:SEQuence2:HYSTeresis:VOLTage
TRIGger:ACQuire:HYSTeresis:VOLTage
This command defines a band around the trigger level through which the signal must pass before an
internal measurement can occur. The band limit above and below the trigger level is one half of the
hysteresis value added to or subtracted from the trigger level.
For a positive trigger to occur, the excursion of an output waveform in the positive direction must start
below the lower hysteresis band limit and pass through the upper hysteresis band limit. For a negative
trigger to occur, the excursion of an output waveform in the negative direction must start above the upper
hysteresis band limit and pass through the lower hysteresis band limit.
Command Syntax
TRIGger:SEQuence2:HYSTeresis:VOLTage<NRf+>
TRIGger:ACQuire:HYSTeresis:VOLTage<NRf+>
Parameters
0 to MAX (see table 8-3)
Unit
V (volts)
*RST Value
0
Examples
TRIG:SEQ2:HYST:VOLT 2
TRIG:ACQ:HYST:VOLT 2
Query Syntax
TRIGger:SEQuence2:HYSTeresis:VOLTage?
TRIGger:ACQuire:HYSTeresis:VOLTage?
Returned Parameters
<NR3>
Related Commands
TRIG:SEQ2:HYST:CURR TRIG:SEQ2:LEV:VOLT

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 66321D and is the answer not in the manual?

Agilent Technologies 66321D Specifications

General IconGeneral
BrandAgilent Technologies
Model66321D
CategoryPower Supply
LanguageEnglish

Related product manuals