EasyManua.ls Logo

Agilent Technologies 6674A

Agilent Technologies 6674A
161 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18 Verification
Table 2-4. Constant Voltage (CV) Tests (continued)
Action
Normal Result
Transient Recovery Time
This test measures the time required for the output voltage to return to within 0.1% or 100mV, whichever is greater, of its
final value following a 50% change in output load current. Measurements are made on both the unloading transient (from
full load to 1/2 load) and the loading transient (from 1/2 load to full load).
1
Turn off the power supply and connect an oscilloscope across +LS and
-LS (see Figure 2-1A).
2 Turn on the power supply and program the current to its maximum
programmable value and the voltage to its full-scale value (see Table 2-2).
3 Program the Electronic Load as follows:
þ Operating mode to constant current.
þ Input load current to 1/2 the supply's full rated output current.
þ Transient current level to the supply's full rated output current.
þ Transient generator frequency = 100Hz.
þ Transient generator duty cycle = 50%.
4 Turn on the transient and adjust the oscilloscope to display response
waveform.
See Figure 2-2.
5 Measure both the loading and unloading transients by triggering the
oscilloscope on both the negative and positive slopes of the transient.
Record the voltage level obtained at the 900-μs interval .
Specified voltage level is reached within
900μs.
Figure 2-2. Transient Response Waveform
100 mV or 0.1%
(see Table 2-2)
100 mV or 0.1%
(see Table 2-2)

Table of Contents

Related product manuals