Step 1: Set Up Mainframe for AC Power
WARNING SHOCK HAZARD. Disconnect power from the mainframe before doing any
installation steps.
A Locate line voltage
selector
115 Vac
Setting
B Is the voltage set correctly?
Yesgo to Step 2 on the next page
Noperform steps C through F below
230 Vac
Setting
C Open door
D Remove selector drum and re-install
with correct voltage facing out
E Install correct fuse for your line voltage
115 Vac = 3.0A fuse
230 Vac = 1.5A fuse
F Close door