For example,CNFG ERROR: 13 shows that the Logical Address Switch may be set wrong. See
Appendix A for the different error messages and possible causes.
Note
If “CNFG ERROR: 3 is noted, download the driver for the module that caused this
error. After the driver is downloaded, perform the above step again. The procedure for
downloading drivers is in the Operating Note that came with the driver.
Checking for System Errors
Send the following SCPI command:
SYSTem:ERRor?
If the mainframe is operational, the command returns this error message:
+0,"No error"
Checking and Setting the System Time
To check the system time, send:
SYSTem:TIME?
The returned message for the time is a comma separated list:
hh,mm,ss
where “hh is the hour,mm are the minutes, and “ss” are the second. For example, if the returned
message is+ 23,+ 5,+ 38”, it shows a time of 23:05:38 or 11:05:39 P.M.. The time is in the 24 hour
format.
To set the time, send:
SYSTem:TIME < hour> ,< minute> ,< second>
For example, to set the time for 10:15:00 AM, send:
SYST:TIME 10,15,00
Checking and Setting the System Date
To check the system date, send:
SYSTem:DATE?