command:
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue:PAIR?
syntax:
:CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:VALue:PAIR?<wsp><index | wavelength[PM | NM |
UM | MM| M],>
description: Queries an offset value pair (wavelength:offset) from the position, or index, of the associated
wavelength in the offset table. Offset table entries are ordered from shortest to longest wave-
length.
Or: Queries the offset value pair (wavelength:offset) applied for the specified wavelength.
parameters: <index>
<wavelength>
The position of the wavelength; offset value pair in the offset table.
The wavelength for the offset table entry, in m
response: char$ in SCPI block format (Intel byte order); wavelength:offset
example: CONF1:OFFS:WAV:VAL:PAIR? 1 "+1.55000000E-006:2"
affects: All attenuator modules.
command:
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue:DELete
syntax:
:CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:VALue:DELete<wsp><index | wavelength[PM | NM
| UM | MM| M],>
description: Deletes an offset value pair (wavelength:offset) from the position, or index, of the associated
wavelength in the offset table. Offset table entries are ordered from shortest to longest wave-
length.
Or: Deletes the offset value pair (wavelength:offset) associated with the specified wavelength.
NOTE Deleting a value pair decrements the index value of every subsequent value pair by 1. When us-
ing this command, you may prefer to work from large to small index values.
parameters: <index>
<wavelength>
The position of the wavelength:offset value pair in the offset table.
The wavelength for the offset table entry, in m
response: none
example: CONF1:OFFS:WAV:VAL:DEL 1
affects: All attenuator modules.
command:
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:VALue:DELete:ALL
syntax: :CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:VALue:DELete:ALL
CAUTION This command clears the offset table!
description: Deletes every value pair (wavelength:offset) from the offset table.
parameters: none
response: none
example: CONF1:OFFS:WAV:VAL:DEL:ALL
affects: All attenuator modules.