EasyManua.ls Logo

Agilent Technologies 8163A - Page 164

Agilent Technologies 8163A
274 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurement Operations & Settings Signal Conditioning
164 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command:
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:TABle?
syntax: :CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:TABle?
description: Queries the complete the offset table.
parameters: none
response: SCPI binary block format format (Intel byte order); wavelength:offset pairs in ascending
order.
Each value pair is transferred as 12 bytes; 8 bytes represent the wavelength, 4 bytes rep-
resent the offset.
example: CONF1:OFFS:WAV:TAB? binary block interpreted as, for example:
1.55e-6 | 12 | 1.7e-6 |3.4 |.....
affects: All attenuator modules.
command:
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:TABle:SIZE?
syntax: :CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:TABle:SIZE?<wsp>MAX | MIN
description: Without optional parameter, queries the size of the offset table.
parameters: MAX
MIN
Queries the maximum size of the offset table.
(available flash memory
1000 entries)
Queries the minimum size of the offset table. (should
0)
response: 4 byte unsigned integer; offset table size.
example: CONF1:OFFS:WAV:TAB:SIZE? 50
affects: All attenuator modules.