2.
YO Dela
y Adjustment
Description
This adjustmen
tuses
the in
ternal coun
ter to
measure
sw
ept
frequency
accuracy
.
YO
dela
y
aects
the
sw
ept
frequency
accuracy
at
fast
sw
eep
times
(less
than
100
ms).
This
pro cedure
adjusts
the
calibration
constants
for the
YIG oscillator
to ac
hiev
e
consisten
t
sw
ept
frequency
accuracy o
v
er
the
en
tire
op
erating
range
of
the
instrumen
t.
CA
UTION
If
y
our
instrumen
t
has
Option
006,
b
e
certain
that
y
ou
iden
tied
y
our
option
conguration
when
the
soft
w
are w
as initially
loaded or
else this
pro
cedure
will
not run
properly
.
Select
the YO
delay
adjustment
and
follo
w
the
prompts
on
the
displa
y
.
No
test
equipment
is required.
5-12 Automated Tests Agilent 8360