EasyManua.ls Logo

Agilent Technologies 8360 B Series - Page 408

Agilent Technologies 8360 B Series
603 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.
ADC Adjustment
Description
This adjustmen
t uses
an external
digital v
oltmeter (D
VM)
to
measure
V
COMP
on
the
YO
driv
er
assem
bly
and
compares
the
reading
to
the
ADC
measuremen
t.
The
ADC
calibration
constan
ts
are
adjusted
so
that
the
DVM
and ADC
measurements
are the
same.
The
pro cedure
is rep
eated for
eac
h
v
oltage
range
(b
oth
plus
and
min
us
supplies).
Select
the ADC
adjustment
and
follo
w
the
prompts
on
the
displa
y
.
The follo
wing test
equipmen
t
is
required
for
this
adjustmen
t.
(The
test
setup
is
giv
en
in
the
soft
w
are.)
Test
Equipment
Required
Instrumen
t
HP/Agilen
t
Mo
del
Num
b
er
Digital V
oltmeter
3456A, 3457A,
or
3458A
Agilent 8360 Automated Tests 5-13