EasyManua.ls Logo

Agilent Technologies 8753ES

Agilent Technologies 8753ES
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1-28
Supplement for the User’s Guide (Firmware Version 7.66)
Conversion Loss Using the Frequency Offset Mode
Performing a Power Meter Calibration Over the RF Range
1. Make the connections as shown in Figure 1-20 (or Figure 1-19) for the one-sweep power
meter calibration over the RF range.
Figure 1-20 8753ET/ES Power Meter Calibration Equipment Setup
for the RF Range
Figure 1-21 8753ES Option 011 Power Meter Calibration Equipment Setup
for the RF Range
2. Use the power meter settings that were set up in steps 1 through 4 of the “Performing a
Power Meter (Source) Calibration Over the IF Range” section.
Perform a one-sweep power meter calibration over the RF frequency range at 0 dBm by
pressing .
NOTE Do not reduce the number of points to perform this power meter calibration.
Reducing the number of points will turn off the receiver calibration.
Cal
PWRMTR CAL
ONE SWEEP
0 x1
TAKE CAL SWEEP

Table of Contents

Other manuals for Agilent Technologies 8753ES

Related product manuals