EasyManuals Logo

Agilent Technologies B1500A Training Manual

Agilent Technologies B1500A
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
Module 7
Measurement Functions
7-42
Calculated value
1.5 M ohm (with Series Resistor)
SMU Series Resistor
Compensation
If you use the SMU series resistor, you will need to compensate the measurement data to eliminate
the effects of the series resistor (1 M ohm).
This is a compensation example by using the following user function.
Vcalc = V1 – I1 * 1000000
The Rcalc value is obtained by using the analysis function and the auto analysis function as same as
the IV-res setup.
This compensation example is shown in the Rcalc setup.

Other manuals for Agilent Technologies B1500A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies B1500A and is the answer not in the manual?

Agilent Technologies B1500A Specifications

General IconGeneral
BrandAgilent Technologies
ModelB1500A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals