Chapter 4 Remote Interface Reference
An Introduction to the SCPI Language
113
4
Command Separators
A colon ( : ) is used to separate a command keyword from a lower-level
keyword as shown below:
"SOURce:CURRent:TRIGgered"
A semicolon ( ; ) is u sed to se pa rat e two com ma n ds wi thin the sam e su bsys tem ,
and can also minimize typing. For example, sending the following command
string:
"SOUR:VOLT MIN;CURR MAX"
... is the same as sending the following two commands:
"SOUR:VOLT MIN"
"SOUR:CURR MAX"
Use a colon and a semicolon to link commands from different subsystems.
For example, in the following command string, an error is generated if you do
not use the colon and semicolon:
‘‘
DISP:TEXT:CLE;:SOUR:CURR MIN
’’
Using the MIN and MAX Parameters
You can substitute
MINimum
or
MAXimum
in place of a parameter for many
commands. For example, consider the following command:
CURRent {<current>|MIN|MAX}
Instead of selecting a specific current, you can substitute
MINimum
to set the
current to its minimum value or
MAXimum
to set the current to its maximum
value.

Other manuals for Agilent Technologies E3632A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3632A and is the answer not in the manual?

Agilent Technologies E3632A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3632A
CategoryPower Supply
LanguageEnglish

Related product manuals