Chapter 5 Error Messages
Execution Errors
123
5
Execution Errors
-101
Invalid character
An invalid character was found in the command string. You may have inserted
a character such as #, $, or % in the command keyword or within a parameter.
Example:
OUTP:STAT #ON
-102
Syntax error
Invalid syntax was found in the command string. You may have inserted a blank
space before or after a colon in the command header, or before a comma.
Example:
VOLT:LEV ,1
-103
Invalid separator
An invalid separator was found in the command string. You may have used a
comma instead of a colon, semicolon, or blank space - or you may have used
a blank space instead of a comma.
Example:
TRIG:SOUR,BUS
or
APPL 1.0 1.0
-104
Data type error
The wrong parameter type was found in the command string. You may have
specified a number where a string was expected, or vice versa.
-105
GET not allowed
A Group Execute Trigger (GET) is not allowed within a command string.
-108
Parameter not allowed
More parameters were received than expected for the command. You may
have entered an extra parameter, or you added a parameter to a command that
does not accept a parameter.
Example:
APPL? 10
-109
Missing parameter
Fewer parameters were received than expected for the command. You omitted
one or more parameters that are required for this command.
Example:
APPL

Other manuals for Agilent Technologies E3632A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3632A and is the answer not in the manual?

Agilent Technologies E3632A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3632A
CategoryPower Supply
LanguageEnglish

Related product manuals