Chapter 3 Front-Panel Operation
Programming Overcurrent Protection
47
3
3
Clear the overcurrent condition and exit this menu.
Now, when you press key again, the “DONE’’ message is
displayed for a second and the
OCP annunciator will not blink any more. The
output will return to meter mode.
Remote interface operation:
CURR:PROT {<current>|MIN|MAX} Set the OCP level
CURR:PROT:STAT {OFF|ON} Disable or enable the OCP circuit
CURR:PROT:CLE Clear the tripped OCP circuit
Over Current