Agilent InfinityLab LC Series Multicolumn Thermostats User Manual 118
7 Error Information
MCT Error Messages
PELTIER OVERCURRENT
Error ID: 32098 (left), 32099 (right)
LEFT PELTIER ERROR
Error ID: 32096
Left Peltier hardware is not operational.
RIGHT PELTIER ERROR
Error ID: 32097
Right Peltier hardware is not operational.
Probable cause Suggested actions
1 Defective sensor
Please contact your Agilent service representa-
tive.
2 Defective mainboard
Please contact your Agilent service representa-
tive.
3 Defective cable
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1 Defect heater element
Please contact your Agilent service representa-
tive.
2 Defective mainboard
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1 Defect heater element
Please contact your Agilent service representa-
tive.
2 Defective mainboard
Please contact your Agilent service representa-
tive.