Electrical Connections
Electrical Connections
• The CAN bus is a serial bus with high-speed data transfer. The two
connectors for the CAN bus are used for internal module data transfer and
synchronization.
• The ERI/REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
start, stop, common shutdown, prepare, and so on.
• With the appropriate software, the LAN connector may be used to control the
module from a computer through a LAN connection. This connector is
activated and can be configured with the configuration switch.
• With the appropriate software, the USB connector may be used to control the
module from a computer through a USB connection.
• The power input socket accepts a line voltage of 100 – 240 VAC ± 10 % with a
line frequency of 50 or 60 Hz. Maximum power consumption varies by
module. There is no voltage selector on your module because the power
supply has wide-ranging capability. There are no externally accessible fuses
because automatic electronic fuses are implemented in the power supply.
NOT E
Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.