EasyManuals Logo
Home>Agilent Technologies>Thermostat>G7116B

Agilent Technologies G7116B User Manual

Agilent Technologies G7116B
203 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
Agilent InfinityLab LC Series Multicolumn Thermostats User Manual 66
3 Using the Module
Installing and Using Column ID Tags
Installing and Using Column ID Tags
Install the Column Identification Option
Parts required p/n Description
5067-5915 Column ID Kit Left
5067-5916 Column ID Kit Right
5067-5917 InfinityLab Column Identification Tag
Preparations
Power off the instrument.
CAUTI ON
Electronic boards and components are sensitive to electrostatic discharge
(ESD).
ESD can damage electronic boards and components.
Be sure to hold the column ID modules by the plastic parts, and do not touch
the electrical components. Do not touch the pins of the flex-board
connector.
NOT E
In the Agilent 1260 Infinity II Multicolumn Thermostat (G7116A), the column ID
module is installed on the left side of the MCT only.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7116B and is the answer not in the manual?

Agilent Technologies G7116B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7116B
CategoryThermostat
LanguageEnglish

Related product manuals