Agilent InfinityLab LC Series Multicolumn Thermostats User Manual 52
3 Using the Module
Exchange a Column
Exchange a Column
Tools required p/n Description
5043-0915 Fitting mounting tool
8710-0510 Open-end wrench 1/4 — 5/16 inch
Parts required p/n Description
For 1290:
5067-5957 InfinityLab Quick Connect Assy ST 0.12 mm x 105 mm
(from heat exchanger to column inlet)
5500-1191 InfinityLab Quick Turn Capillary ST 0.12 mm x 280 mm, long socket
(from column to detector)
5067-5966 InfinityLab Quick Turn Fitting
For 1260:
5500-1193 InfinityLab Quick Turn Capillary ST 0.17 mm x 105 mm, long socket
(from heat exchanger to column)
5500-1191 InfinityLab Quick Turn Capillary ST 0.12 mm x 280 mm, long socket
(from column to detector)
5067-5966 InfinityLab Quick Turn Fitting
(column inlet)
0100-1516 Finger-tight fitting PEEK, 2/pk
(column outlet)
Bio-inert option:
5067-4780 Bio Capillary 0.17 mm x 300 mm, 1.6 mm OD socket
(from column to detector)
5067-5966 InfinityLab Quick Turn Fitting
(column inlet)
0100-1516 Finger-tight fitting PEEK, 2/pk
(column outlet)
Bio-compatible option:
5067-5966 InfinityLab Quick Turn Fitting
5500-1596 Quick Turn Capillary MP35N 0.12 mm x 280 mm
5500-1578 Quick Connect Capillary MP35N 0.12 mm x 105 mm
5067-5965 InfinityLab Quick Connect LC fitting
Preparations
• Heat exchanger(s) installed in MCT
• Column holder clip(s) installed in the heater block on top of heat exchanger(s)
• Column inlet capillary connected to heat exchanger outlet