Agilent InfinityLab LC Series Diode Array Detectors User Manual 118
7 Error Information
Detector Error Messages
UV Heater Current
Error ID: 7453
The UV lamp heater current is missing.
During UV lamp ignition, the processor monitors the heater current. If the current
does not rise above the lower limit within one second, the error message is
generated.
Calibration Values Invalid
Error ID: 1036
The calibration values read from the spectrometer ROM are invalid.
After recalibration, the calibration values are stored in ROM. The processor
periodically checks if the calibration data are valid. If the data are invalid or
cannot be read from the spectrometer ROM, the error message is generated.
Probable cause Suggested actions
1 Lamp disconnected.
Ensure the UV lamp is connected.
2 Ignition started without the top foam in
place.
Please contact your Agilent service representa-
tive.
3 Defective UV lamp or non-Agilent lamp.
Exchange the UV lamp.
4 Defective detector main board.
Please contact your Agilent service representa-
tive.
5 Defective power supply.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1 Defective connector or cable.
Please contact your Agilent service representa-
tive.
2 Defective PDA/optical unit.
Please contact your Agilent service representa-
tive.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117A and is the answer not in the manual?

Agilent Technologies G7117A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117A
CategorySecurity Sensors
LanguageEnglish

Related product manuals