EasyManuals Logo

Agilent Technologies G7117A User Manual

Agilent Technologies G7117A
230 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #149 background imageLoading...
Page #149 background image
Agilent InfinityLab LC Series Diode Array Detectors User Manual 149
8 Test Functions and Calibration
D/A Converter (DAC) Test
Figure 48 D/A Converter (DAC) Test – Example of Integrator Plot
Test Evaluation
The noise on the step should be less than 3 µV.
Probable cause Suggested actions
1 Bad cable or grounding problem between
detector and external device.
Check or replace the cable.
2 Defective detector main board.
Please contact your Agilent service representa-
tive.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117A and is the answer not in the manual?

Agilent Technologies G7117A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117A
CategorySecurity Sensors
LanguageEnglish

Related product manuals