Agilent InfinityLab LC Series Diode Array Detectors User Manual 136
8 Test Functions and Calibration
Quick Noise Test
Test Failed
Quick Noise Test Evaluation
Probable cause Suggested actions
1 Insufficient lamp warm-up time.
Allow detector and UV-lamp turned on for at least
2 hours.
2 Absorbing solvent or air bubble in flow cell.
Ensure the flow cell is filled with water, and free
from air bubbles.
3 Dirty or contaminated flow cell.
Flush flow cell.
Clean the flow cell as described in “Clean the
Max-Light Cartridge Cell” on page 165.
4 Old UV-lamp.
Exchange the UV-lamp.