Agilent InfinityLab LC Series Diode Array Detectors User Manual 166
9 Maintenance
Storage of Max-Light Cartridge Cell
Storage of Max-Light Cartridge Cell
1 Flush the Max- Light Cartridge Flow Cell with iso- propanol or methanol and
insert the plugs into the cell inlet and outlet (see “Replace the Max-Light
Cartridge Cell” on page 162).
2 Remove the Max-Light Cartridge Cell from the cartridge holder of the detector.
3 Replace the black hoods, that secure the cell light inlet and outlet.
4 Store the Max-Light Cartridge Cell in plastic case provided with the Max-Light
Cartridge Flow Cell.