Getting started
6 G5585B PlateLocThermal Microplate Sealer Quick Guide
Checking the internal air pressure
To check the internal air pressure:
On the touchscreen Home page, locate the Air Pressure readout. The regulated air
pressure should be approximately 0.599–0.62
MPa (87–90 psi).
To turn off the air:
Push the AIR ON/OFF switch to the OFF position.
Loading and unloading a roll of seal
Make sure you have the supplied seal-loading card.
To warm up or cool down the PlateLoc Sealer:
1 On the touchscreen, press to display the Settings pages.
2 Set the Seal Temp to a value from 30 to 40 °C.
Note: Press or to adjust the value down or up in 0.5 °C increments.
3 Press Save to save the new Seal Temp value.