XGS-600 Gauge Controller
DRAFT 4/15/14
45
Mains Fuse Replacement Procedure
WARNING Only trained service personnel should attempt this work.
Disconnect the XGS-600 mains connection by unplugging the
IEC power cord from the back of the unit before removing the
unit's cover.
WARNING Replace the mains fuse only with the same type and electrical
ratings (250V, 5A, FAST-BLO, 5x20mm).
1. Turn off and unplug the line cord (AC mains).
2. Remove the cover by unscrewing the two Philips head screws.
3. Locate the mains fuse under the IEC input module and replace (Figure 4-5).
Figure 4-5 Fuse Replacement
4. Replace the cover by:
a. Pushing the cover straight in under the lip at the XGS front bezel.
b. Replacing the two Philips head screws.
c. Ensuring the cover is flat and fully seated. It may be necessary to adjust the
boards and filler plates slightly for the cover to fully engage. The cover should
drop on front of the filler plate.
5. Plug the unit back in and turn on the power switch.