Parametri per il funzionamento della colonna:
pH e temperatura
Fase Parametro pH
consigliato
Max temperatura
operativa
InfinityLab Poroshell 120
EC-C18, EC-C8 e Bonus-RP.
ZORBAX Eclipse Plus C18 e C8.
ZORBAX Eclipse XDB-C18,
XDB-C8 e XDB-Phenyl ZORBAX
Bonus-RP.
HC-C18(2) e TC-C18(2).
da pH 2,0 a 9,0 60 °C
InfinityLab Poroshell 120 PFP e
Phenyl-Hexyl.
ZORBAX Eclipse Plus
Phenyl-Hexyl.
ZORBAX XDB-CN.
ZORBAX Eclipse PAH.
ZORBAX ODS e Phenyl.
Tutte le fasi legate Pursuit,
Polaris, Microsorb, Metasil,
Monochrom, Hypersil,
OmniSpher, MicroSpher.
da pH 2,0 a 8,0 60 °C
ZORBAX Eclipse PAH da pH 2,0 a 8,0 60 °C a pH <6,0
ZORBAX TMS da pH 2,0 a 7,0 60 °C
InfinityLab Poroshell 120
SB-C18, SB-C8 e SB-Aq.
ZORBAX StableBond SB-C18,
SB-C8, SB-Aq, SB-C3 e
SB-Phenyl
ZORBAX Rx-C8.
da pH 1,0 a 8,0 90 °C (StableBond C18 e
Poroshell 120 SB-C18)
80 °C (Rx-C8 e SB-Aq,
SB-C3 e SB-Phenyl)
InfinityLab Poroshell HPH-C18
e HPH-C8
da pH 3,0 a 11,0 60 °C
ZORBAX Extend C18 da pH 2,0 a 11,5 60 °C (40 °C a pH elevato)
Note: tutti i riempimenti in silice sono solubili in fasi mobili acquose a pH >6. Se si
utilizzano colonne in silice a pH >6, è preferibile impostare temperature più basse
(40°C max) per prolungare la durata della colonna e utilizzare basse concentrazioni
di tamponi (da 0,01 a 0,02 M). L’utilizzo agli estremi dell’intervallo consentito di pH e
temperatura può avere un impatto significativo sulla durata di vita delle colonne.