Page #5 background image
Column Operating Parameters: pH and Temperature
Phase Recommended
pH range
Maximum operating
temperature
InfinityLab Poroshell 120 EC-C18,
EC-C8 and Bonus-RP.
ZORBAX Eclipse Plus C18 and C8.
ZORBAX Eclipse XDB-C18,
XDB-C8 and XDB-Phenyl ZORBAX
Bonus-RP.
HC-C18(2) and TC-C18(2).
pH 2.0 to 9.0 60 °C
InfinityLab Poroshell 120 PFP and
Phenyl-Hexyl.
ZORBAX Eclipse Plus Phenyl-Hexyl.
ZORBAX XDB-CN.
ZORBAX Eclipse PAH.
ZORBAX ODS and Phenyl.
All Pursuit, Polaris, Microsorb,
Metasil, Monochrom, Hypersil,
OmniSpher, MicroSpher bonded
phases.
pH 2.0 to 8.0 60 °C
ZORBAX Eclipse PAH pH 2.0 to 8.0 60 °C at pH <6.0
ZORBAX TMS pH 2.0 to 7.0 60 °C
InfinityLab Poroshell 120 SB-C18,
SB-C8 and SB-Aq.
ZORBAX StableBond SB-C18, SB-
C8, SB-Aq, SB-C3 and SB-Phenyl
ZORBAX Rx-C8.
pH 1.0 to 8.0 90 °C (StableBond C18
and Poroshell 120 SB-C18)
80 °C (Rx-C8 and SB-Aq,
SB-C3 and SB-Phenyl)
InfinityLab Poroshell HPH-C18 and
HPH-C8
pH 3.0 to 11.0 60 °C
ZORBAX Extend C18 pH 2.0 to 11.5 60 °C (40 °C at high pH)
Note: All silica-based packings have some solubility in pH >6 aqueous mobile
phases. When using silica-based columns at pH >6, best column lifetime is
obtained at lower temperatures (40 °C max) using low buffer concentrations in the
range of 0.01 to 0.02 M. Operating at extreme ends of pH and temperature ranges
will have a significant impact on column lifetime.
ENGLISH

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies ZORBAX Series and is the answer not in the manual?

Agilent Technologies ZORBAX Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelZORBAX Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals