PORTUGUÊS
Parâmetros operacionais da coluna: pH e temperatura
Fase Faixa de pH
recomendada
Temperatura
operacional máxima
InfinityLab Poroshell 120 EC-C18,
EC-C8 e Bonus-RP.
ZORBAX Eclipse Plus C18 e C8.
ZORBAX Eclipse XDB-C18, XDB-C8 e
XDB-Phenyl ZORBAX Bonus-RP.
HC-C18(2) e TC-C18(2).
pH 2,0 a 9,0 60 °C
InfinityLab Poroshell 120 PFP
e Phenyl-Hexyl.
ZORBAX Eclipse Plus Phenyl-Hexyl
ZORBAX XDB-CN.
Eclipse ZORBAX PAH.
ZORBAX ODS e Phenyl.
Fases ligadas All Pursuit, Polaris,
Microsorb, Metasil, Monochrom,
Hypersil, OmniSpher e MicroSpher.
pH 2,0 a 8,0 60 °C
ZORBAX Eclipse PAH pH 2,0 a 8,0 60 °C a pH <6,0
ZORBAX TMS pH 2,0 a 7,0 60 °C
InfinityLab Poroshell 120 SB-C18,
SB-C8 e SB-Aq.
ZORBAX StableBond SB-C18, SB-C8,
SB-Aq, SB-C3 e SB-Phenyl
ZORBAX Rx-C8.
pH 1,0 a 8,0 90°C (StableBond C18
e Poroshell 120 SB-C18)
80°C (Rx-C8 e SB-Aq, SB-C3
e SB-Phenyl)
InfinityLab Poroshell HPH-C18
e HPH-C8
pH 3,0 a 11,0 60 °C
ZORBAX Extend C18 pH 2,0 a 11,5 60°C (40°C em pH elevado)
Observação: todos os empacotamentos à base de sílica têm alguma solubilidade
em fases móveis aquosas com pH maior que 6. Ao utilizar colunas à base de sílica
em pH maior que 6, uma melhor vida útil da coluna é obtida em temperaturas mais
baixas (máx. 40°C) usando concentrações baixas para o tampão na faixa
de 0,01 a 0,02M. Ao operar nos extremos das faixas de pH e temperatura causará
um impacto significativo sobre a vida útil da coluna.