Troubleshooting
TROUBLESHOOTING THE BALER HAY TESTER
PROBLEM: Tester will not power on.
SOLUTION:
• Check all power connections and the 2 amp fuse inside the fuse holder.
• Use a test light to ensure that the power cable has power.
PROBLEM: Tester will not calibrate and displays “Err” during calibration.
SOLUTION:
• Check all cable and sensor connections to ensure nothing is shorted.
• Make sure the chamber is empty before performing the calibration.
• With a light steel wool pad, polish both sensor pads and retry calibration.
PROBLEM: Tester only displays “HI” for the moisture reading.
SOLUTION:
• Check all cable and sensor connections to ensure nothing is shorted.
• Determine by some other means if the hay being tested is with the
operating limits of the tester.
PROBLEM: Tester only displays “LO” for the moisture reading.
SOLUTION:
• Check all cable and sensor connections to ensure nothing is open.
• Determine by some other means if the hay being tested is with the
operating limits of the tester.
PROBLEM: Tester reading seems to read erratic or incorrect.
SOLUTION:
• See the section of the manual “UNDERSTANDING HAY CONDITIONS
AND TEST READINGS”.
If all of these steps fail, please read the manual again carefully or contact
customer service using the contact information provided in this manual.