-23- -24-
Whenperformingmeasurementwhichwilldestroyor
cutanobjecttomeasure,wearequipmenttoprotect
yourselffrombrokenpieceshittingyoureyesorbody.
Measurementofswitch/relaycontactforce
■Printer
Externalholdfunction
Aforcewhenacontact
isONorOFFcanbe
measured.
※Formeasurement,use
theoptionalcable
RZ-OP-2.
※No-voltagecontact
input.
LineThermalPrinterBL2-58SNWJCor
DigimaticMiniProcessorDP-1VR
Howtousetheexternalcontactholdfunction
【Example1】Switchoperatingforcemeasurement
Anexampleofcombinationwithanelectricstand
MODEL-1308U
Capacity/1000N
BL2−58SNWJC
BL2−58SNWJCPrintingsample
0001+4.65 N
−7.73 N
0002+8.53 N
−6.85 N
0003+0.03 N
−0.02 N
0004+0.08 N
−1.40 N
0005+0.34 N
−2.39 N
Measurement②
<Measurementbyuseoftheexternalcontacthold>
①ConnectingtheRZgageandaprinter
ConnecttheI/OconnectoratthebottomoftheRZgageandaprinterwiththe
optionalcable.ToconnecttoBL2-58SNWJC,usetheoptionalcableRZ-OP-4
andtoconnecttoDP-1VR,usetheoptionalcableRZ-OP-3.
②Connectingtheexternalcontact
Theexternalcontactholdsignallineshavebeenprovidedwhentheoptional
cableisused.TheyarePin14(HoldIn)andPin17(GND)ofthe26-pin
connectoratthebottomofthegage.
(Thecontinuousshort-circuit/releasevaluesofOFF→ON→OFFareheldand
indicated.)
<Measurementprocedurebyuseoftheexternalcontactholdfunction>
Contactinputdiagram
※Fortheprocedurethatfollows,seeMeasurement①
 <Measurementbyuseofthememoryfunction>.
HONDATSUSHIN
KOGYO-made
26-pinconnector

Related product manuals