64 AS-D+—AS-P Series Instruction Manual
AirSep Corporation
High bed pressure alarm 83 psig
High bed pressure shutdown 85 psig
Low oxygen concentration alarm 89%
Low oxygen concentration shutdown setpoint 87%
Time delay for concentration shutdown at
start-up
1 hour
Cycle pressure/feed air regulator setting 70-75 psig
Oxygen sample line pressure regulator setting 1 psig
Table A.9: Typical System Set points