Questions and Answers

  • R
    Ruth LewisSep 10, 2025
    What to do if AirSep Medical Equipment FreeStyle has rapid alarm condition, and the yellow BATT light illuminates intermittently, and FreeStyle shuts down?
    • C
      Christopher LambertSep 10, 2025
      If your AirSep Medical Equipment FreeStyle is experiencing a rapid alarm, the yellow BATT light is flashing, and the unit shuts down, it indicates that the battery voltage is too low. Immediately connect the unit to a DC or AC outlet, or use the optional AirBelt.
  • J
    Joyce Bryant MDSep 9, 2025
    Why does my AirSep Medical Equipment FreeStyle have intermittent alarm condition, and the yellow batt light illuminates intermittently?
    • S
      Sierra AcostaSep 9, 2025
      If your AirSep Medical Equipment FreeStyle has an intermittent alarm and the yellow BATT light is flashing, it means the battery needs charging. Connect the unit to a DC or AC outlet within 2 minutes, or connect it to the optional AirBelt.
  • M
    matthewericksonSep 7, 2025
    What to do if AirSep Medical Equipment FreeStyle has intermittent alarm condition, and the red alarm light illuminates intermittently?
    • P
      Philip BauerSep 7, 2025
      If your AirSep Medical Equipment FreeStyle has an intermittent alarm and the red alarm light flashes, it means that your breathing rate has exceeded the capacity of the unit. In this case, switch to another source of oxygen and contact your Equipment Provider.
  • J
    Jennifer JonesAug 22, 2025
    What to do if AirSep Medical Equipment has continuous alarm?
    • P
      paul20Aug 23, 2025
      If your AirSep Medical Equipment FreeStyle unit sounds a continuous alarm and the red alarm light stays on, it means the device hasn't detected a breath for 15 minutes. Check the cannula connection, make sure the tubing isn't kinked, and ensure the cannula is properly positioned so you're breathing through your nose. If you breathe through your mouth, a chin strap might help.
  • K
    Kenneth WestSep 5, 2025
    How to fix AirSep Medical Equipment FreeStyle that does not operate when a pulse flow selection button is pressed?
    • D
      Donald GeorgeSep 5, 2025
      If your AirSep Medical Equipment FreeStyle unit does not operate when you press a pulse flow selection button, the battery might be discharged. Try powering the unit through the DC outlet or an AC outlet. If this doesn't resolve the issue, it may indicate a malfunction, and you should contact your Equipment Provider and switch to a reserve oxygen supply.
  • M
    mwhiteSep 4, 2025
    What causes delay in recharging internal battery of AirSep?
    • N
      Nicholas RiveraSep 4, 2025
      A delay in recharging the internal battery of your AirSep unit can occur if the internal battery exceeds its charging temperature. The unit can still be operated, but charging may not resume until the battery temperature decreases.
  • D
    Diana WolfSep 3, 2025
    How to troubleshoot AirSep FreeStyle if it does not operate?
    • J
      jturnerSep 3, 2025
      If your AirSep FreeStyle doesn't operate when you press a pulse flow selection button, it could be due to a discharged battery or a malfunction. Try powering the unit through a DC or AC outlet.
  • D
    Douglas WoodsSep 2, 2025
    What to do if AirSep FreeStyle has an intermittent alarm and yellow light?
    • S
      Steven CharlesSep 2, 2025
      If your AirSep FreeStyle has an intermittent alarm and the yellow BATT light flashes, it means the battery needs charging. Connect it to a DC or AC outlet within 2 minutes, or connect to the optional AirBelt.
  • O
    ofullerSep 1, 2025
    What to do if AirSep FreeStyle has an intermittent alarm and red light?
    • S
      Savannah WilliamsSep 1, 2025
      If your AirSep FreeStyle has an intermittent alarm and the red light flashes, it means your breathing rate has exceeded the unit's capacity. Reduce your activity, then turn the unit off and back on to reset it.
  • A
    adamslauraAug 31, 2025
    Why AirSep FreeStyle won't start on battery power?
    • R
      Rachel ValenzuelaAug 31, 2025
      If your AirSep FreeStyle won't start on battery power, it might be too hot or cold from being left outdoors, like in a car. Let it reach normal operating temperature, which could take a few minutes if it was exposed to extreme temperatures. You may temporarily connect your AC or DC power supply to the unit’s power inlet and power source, as needed, to reset the unit’s internal battery.

Summary

Introduction to FreeStyle Service

Equipment Provider Responsibility

Outlines the duties of equipment providers for handling, operation, and maintenance.

Important Notices and Symbol Explanations

Explains safety symbols (Warning, Caution, Note) and their meanings.

Definitions of Symbols

Lists and defines various symbols used in the manual and on the equipment.

FreeStyle Product Limited Warranty

Details the warranty terms and conditions for the FreeStyle Oxygen Concentrator.

FreeStyle Specifications

Provides technical specifications for the FreeStyle unit, including dimensions, power, and performance.

FreeStyle Classifications

Details the electrical and safety classifications of the FreeStyle equipment.

Operation Check and Oxygen Concentration Test

Description of Operation

Explains how the FreeStyle oxygen concentrator works to produce oxygen.

Operation Check Procedure

Procedure to test the FreeStyle unit after manufacture for shipping damage.

Alarm System Overview

Covers startup, battery status, low/discharged battery, charging, full charge, disconnect, performance, capacity, and general malfunction alarms.

Oxygen Concentration Test and Specification

Procedure to test and verify the oxygen output concentration of the FreeStyle unit.

Patient Instructions and Maintenance

General Operation and Power

Guidance on unit operation, connecting power, and using the optional AirBelt battery.

Patient Maintenance Procedures

Routine maintenance tasks for patients, including filter cleaning and indicator checks.

Equipment Provider Maintenance

Maintenance Procedures Overview

Overview of maintenance, including filter care, filter replacement, compressor servicing, and record keeping.

Infection Control and Cleaning

Procedures for infection control and cleaning the unit for new patient use.

Service Procedures for FreeStyle

Components and Housing

Overview of components and instructions for removing unit housing.

Internal Battery Replacement

Step-by-step guide for replacing the internal battery pack.

PSA System Service

Procedure for safely removing the PSA system assembly from the unit.

Compressor Replacement

Detailed steps for removing and installing the compressor assembly.

Manifold, Valve, and Bed Replacements

Instructions for replacing manifolds, solenoid valves, and adsorbent beds.

System Pressure Measurement

Method for measuring system pressure using a test kit.

Hour Meter and Controls

Explains hour meter functionality, replacement, and related indicators.

Electronic Component Replacements

Procedures for replacing circuit boards, control panels, and the fan.

FreeStyle Pressure Profiles

Pressure Profile Quadrant

Illustrates normal operating pressure ranges and fault conditions for the unit.

Troubleshooting the FreeStyle Unit

General Troubleshooting Steps

Initial steps to isolate malfunctions before consulting the troubleshooting chart.

Troubleshooting Chart

A detailed chart listing problems, possible causes, and solutions for unit malfunctions.

Recommended Tools for Service

Tools for FreeStyle Service and Maintenance

Lists essential tools for performing service and maintenance on the FreeStyle unit.

Appendix: Exploded Drawings

Exploded Drawings Overview

Provides diagrams showing the internal components of the unit.

AirSep FreeStyle Specifications

General IconGeneral
Audible AlarmYes
Operating Temperature41°F to 104°F (5°C to 40°C)
Power100-240 VAC, 50/60 Hz
Operating Humidity15% to 95% (non-condensing)
Oxygen Concentration90% (+5.5% / -3%) at all settings
Weight4.8 lbs (2.2 kg) with battery