S SERIES CLEAN BENCH
ModelsVS-840KVS-840K-UVS-1300LVS-1300L-U
HS-840HS-840-UHS-1300HS-1300-U
OPERATION, MAINTENANCE & SERVICE
INSTRUCTION

Questions and Answers

  • P
    Pam ScottAug 11, 2025
    What to do if the AirTech Laboratory Equipment fan is in high speed but the air velocity is low?
    • H
      harold06Aug 11, 2025
      If the fan in your AirTech Laboratory Equipment is running at high speed, but the air velocity remains low, it could be due to a dusty pre-filter or HEPA filter failure. Try cleaning or replacing the pre-filter. If the issue persists, replace the HEPA filter.
  • D
    David FlemingAug 17, 2025
    How to fix inoperative buttons on AirTech S Series control panel?
    • J
      Jacob EdwardsAug 17, 2025
      If the buttons on your AirTech Laboratory Equipment control panel are not working, the issue could be due to a lack of power connection or a faulty circuit board/control panel. Ensure the equipment is reliably connected to the power supply. If the problem continues, inspect the wiring and fuse, replacing any faulty components as needed.
  • A
    April StuartAug 23, 2025
    What to do if the fan does not work in AirTech Laboratory Equipment?
    • B
      Bryan SmithAug 23, 2025
      If the fan in your AirTech Laboratory Equipment is not working, the issue could stem from a circuit failure (no output) or a fan failure. Inspect the wiring; if necessary, replace the circuit board. If the wiring is intact, check the fan itself and replace it if needed.
  • M
    Melanie JohnsonAug 28, 2025
    Why the light or UV light cannot be turned on in AirTech Laboratory Equipment?
    • K
      Kevin CampbellAug 28, 2025
      If the light or UV light in your AirTech Laboratory Equipment fails to turn on, several factors could be responsible. A loose or damaged light base or starter, a damaged tube, or a damaged ballast. Begin by replacing the starter. If the issue remains, inspect the tube and replace if damaged. If the problem persists, check the ballast and replace if necessary. Also, check the wiring and replace any faulty parts.
  • W
    williamssydneySep 4, 2025
    What to do if there is no output of the spare socket in AirTech S Series?
    • M
      Michelle PetersonSep 4, 2025
      If there is no output from the spare socket of your AirTech Laboratory Equipment, it is likely due to fusing caused by overloading. Replace the fuse with one of the same capacity and ensure you are using a proper external device that does not overload the socket.