Frequently Asked Questions
Q. How can I clean the pump?
A. With warm, soapy, water dampen a cloth or sponge. Clean the pump housing and rotor.
Clean the drop sensor with soft cotton swabs dipped in isopropyl alcohol. For direction
regarding contact with biohazardous material and cleaning methods not recommended refer
to the SENTINELplus Operator Manual. Follow other official procedures that may differ
among institutions.
Q. What are the minimum and maximum flow rates?
A. 5-295mL/hr in 1 mL increments.
Q. What is the weight and size of the pump?
A. 1.42kg (3.13lbs), 15cm H x 20cm W x 9cm D (5.91in H x 7.87in W x 3.54in D).
Q. What kind of adminstration sets can I use with the SENTINELplus?
A. Use only SENTINELplus administration sets. Use of enteral administration sets other than
SENTINELplus sets will result in an alarm.
Q. What is the battery life of the pump?
A. Battery backup 6 hours minimum at a rate of 125mL/hr.
Q. Does the pump require routine calibration?
A. No, the SENTINELplus does not require any routine calibration or adjustment procedures
with its use. All functions of the formula delivery rate and alarms are software controlled and
not adjustable through any service or normal activity.
Q. Can anyone service and repair the pump?
A. No, there are no user serviceable parts on the pump. Do not remove the back cover.
Service may only be performed by qualified personnel only (ALCOR authorized service
providers).