Alarms and Indicators
DOSE COMP
OCC/EMP
BATTERY INDICATOR
AC POWER INDICATOR
DOSE COMPLETE
This indicator flashes and pump stops when
the quantity of formula delivered reaches
the dose limit setting. There will be two
beeps with a flashing indicator. This is
repeated very 2.5 minutes until action is
taken.
OCCLUSION/EMPTY ALARM
When the pump is running, this alarm indicator
is activated and pump stops if:
o The feeding container is empty
o An occlusion has stopped delivery
o The pump is left on SToP for more than
2.5 minutes
o The drip chamber is installed improperly
BATTERY INDICATOR
When operating on battery power, this
indicator shows solid yellow. Approximately
15 minutes before the battery is discharged,
the light will flash, an alarm will sound. The
pump will not restart if stopped at this time.
AC POWER INDICATOR
When plugged into AC power, this indicator
always shows solid green indicating the battery
is charging. The light remains on even when
pump is turned off. The ALCOR power supply
should always be plugged into an AC outlet
during normal use.
EXCESS FLOW
The displays flash the text FLO FAST and a
three-tone beep sounds whenever liquid
flow rate exceeds preset limits. This can
occur if the set is dislodged from the pump
during operation or the drip sensor is
blocked for more than 6 seconds. The alarm
sounds at maximum volume and the pump
stops.
12