47
USE of Alicat instruments (M, MC, MS, MCS, MCR, P, PC and PCR product families only) in
Class 1 Division 2 applications.
CSA certifi es the use of this product for general use as well as use in hazardous locations as
defi ned by Class 1 Division 2 Group A, B, C and D T4.
CSA certifi cation is indicated by the product label as shown below and not by the
statements in this, or any accompanying documentation.
Special Conditions:
To comply with CSA certifi cation the following information is included in the product literature:
When equipment is properly labeled, it is suitable in Class I, Division 2, Group A, B, C and D, T4•
Tamb. -40°C to +50°C o
Electrical Rating 24Vdc, 0.800A max•
Instruments shall be powered by a CSA certifi ed, UL listed, Class II external power supply suitable for •
the application
Instruments shall be housed in an enclosure with a minimum IP54 rating or location providing equivalent •
protection
Instrument’s fi nal approval shall be provided by the local authority having jurisdiction•
EEx nA IIC T4
Class I, Div. 2 Group A, B, C and D T4
24 Vdc, 0.800A max
Class I, Zone 2 AEx nA IIC T4
X – See manual for special conditions
WARNINGS:
EXPLOSION HAZARD – DO NOT
DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-
HAZARDOUS.
EXPLOSION HAZARD – SUBSTITUTION
OF COMPONENTS MAY IMPAIR
SUITABILITY FOR CLASS I, DIVISION 2.
www.AlicatScientifi c.com
Alicat Scientifi c, Inc.
Tucson, AZ USA
Tel: 520-290-6060
II 3 G
USE of Alicat instruments (M, MC, MS, MCS, MCR, P, PC and PCR product families only) in
applications requiring ATEX Certifi cation.
Properly labeled Alicat instruments comply to the following ATEX standard:
II 3 G EEx nA IIC T4 (-40°C ≤ Ta ≤ +50°C)
The examination certifi cate was issued by the CSA in accordance with accepted practices and
procedures. This confi rms compliance with the European ATEX Directive or Group II Category
3G equipment.
ATEX certifi cation is indicated by the product label as shown above and not by the statements in this, or
any accompanying documentation.
Special Conditions:
Properly labeled equipment is only certifi ed for use in ambient temperatures in the range of -40°C to •
+50°C only
Electrical Rating 24Vdc, 0.800A max•
Instruments shall be powered by a CSA certifi ed, UL listed, Class II external power supply suitable for •
the application
Instruments shall be housed in an enclosure with a minimum IP54 rating or location providing equivalent •
protection
Instrument’s fi nal approval shall be provided by the local authority having jurisdiction•