46
Additional Information for Alicat CSA and ATEX Approved Devices
See the following page for Special Conditions regarding the use of these units!
1
6
5
4
3
2
Pin Function Cable Color
1
Power In (
+ )
Red
2 RS-232TX / RS-485 + Blue
3 RS-232RX / RS-485 – White
4 Analog Input Signal Green
5 Ground (common for power,
communications and signals)
Black
6 Signal Out (Voltage or Current as ordered) Brown
EEx nA IIC T4
Class I, Div. 2 Group A, B, C and D T4
24 Vdc, 0.800A max
Class I, Zone 2 AEx nA IIC T4
WARNINGS:
EXPLOSION HAZARD – DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
EXPLOSION HAZARD – SUBSTITUTION OF COMPONENTS MAY IMPAIR
SUITABILITY FOR CLASS I, DIVISION 2.
II 3 G
Pin# Wire color
1 Red
2 Blue
3 White
4 Green
5 Black
6 Brown
IC-10 Cable
Ø .710
.425
Industrial Connector Overall Clearance
Minimum
Removal
Clearance
Min. Clearance w/
Cable Bend
2.537
1.817
1.602
All Alicat CSA / ATEX approved devices are equipped with a locking 6 pin industrial connector. The
power and signal connections are shown below.
IC-10 Locking Industrial Cable
Clearance Requirements for