Calibrating Your Module
Publication
1794-6.5.7 – April 1997
Gain Calibration
Make sure that you have calibrated the offset for this channel before
calibrating the gain.
1. Connect 75.000mV across each input channel. Connect all high
signal terminals together and attach to the positive lead from the
precision voltage source. Connect all low signal terminals
together and attach to the negative lead.
2. Click on the channels you want to calibrate.
3. Click on the radio button for gain calibration. Then click on
.
4. When calibration is complete, a notification will appear on the
calibration status line.